Operation
1. Program P1/P2: Press P1/P2 key to enter the fixed program, press P1/P2 key for 5 seconds to enter the
program setting interface, press the P1/P2 key to switch the spinning speed, time, centrifugal force, and
adjust the parameters by rotating the knob. The parameters are automatically saved after the setting is
finished for 7 seconds; the parameters of the setting can be memorized and saved, and automatically
recalled when the machine is turned on again;
2. Set operating parameters: Press the knob to select the parameter Speed, RCF or time and rotate to input
or modify accordingly.
3. Set the rotation speed:Press the Select key to choose the rotation speed, adjust the parameter value
in rpm. The minimum rotation speed is 300rpm, the minimum increment is 10rpm. Set the run time:
Set the time parameter value to desired setting within the range of 30s-99min. When the time displays
HD, this is a continuous running mode.
4. Switch the deceleration speed: Press the knob for 5 seconds and after sound reminder, switch between
1 and 2 steps;
Start Run
1)Press Start button to start running
2)View and modify the operation programs
Operation programs can be modified after the centrifuge reaches the set speed. Pressing the knob to return
the program interface and displays setting programs. Press the knob and when parameter flashing, rotate
knob to modify values. Release the button after 5 seconds, and the centrifuge will return to normal operation
mode and run according to the new value.
Stop running
When the set run time is completed the centrifuge stops its run operation automatically or it can be stopped
by pressing the
. Open the outer lid to remove the samples and rotor.
Open the door manually
If power is off by accident, the door can be opened manually by sliding the switch to right which is at the
bottom right front side.