FIRST USE
1. Turn on the Aadva Lab Scan via the ON/OFF switch on the back panel.
The ventilation fans should turn on and the GC logo on the front panel will light up.
After a few seconds, the blue light from the LED in the sensor head will be visible.
Switch on the PC.
2. Once the workstation and monitor are switched on, check to see if the LEDs on
the license dongles blink a few times.
3. Start the Dental manager software via the application shortcut on the desktop.
4. Once the Aadva Lab Scan and the scanning software are running correctly,
the Aadva Lab Scan must be calibrated again. This may be necessary if the Aadva
Lab Scan has been transported in a horizontal position or if the scanner has undergone
rough transport. The Aadva Lab Scan may have to be recalibrated even if the scan
software has not indicated that calibration is necessary.
5. Make sure the calibration plate is clean before starting the different calibration
steps. Use a dry and soft tissue to clean.
Aadva LAB SCAN from GC
12 12