5. Technical Data
ErgoOne
®
Single-Channel
0.1 – 2.5 µl
0.5 – 10 µl
2 – 20 µl (micro cone)
2 – 20 µl (standard cone)
10 – 100 µl
20 – 200 µl
100 – 1,000 µl
500 – 5,000 µl
ErgoOne
®
Multi-Channel
0.5 – 10 µl
10 – 100 µl
30 – 300 µl
Measurement data determined in compliance with the test conditions for piston-stroke pipettes
10
· ErgoOne · User Manual
Technical Data ErgoOne
Pipette
volume [µl]
0.1
0.25
1.25
2.5
1.0
5
10
2
10
20
2
10
20
10
50
100
20
100
200
100
500
1,000
500
2,500
5,000
Pipette
volume [µl]
1.0
5
10
10
50
100
30
150
300
according to EN ISO 8655
Measurement carried out with original TipOne
Pipettes
®
Systematic measure-
ment deviation
(inaccuracy) ± [%]
EN ISO
ErgoOne
®
8655-2
± 125
± 24
± 50
± 12
± 10
± 2.5
± 5.0
± 1.4
± 12
± 2.5
± 2.4
± 1.5
± 1.2
± 1.0
± 10
± 5.0
± 2.0
± 1.2
± 1.0
± 1.0
± 10
± 5.0
± 2.0
± 1.2
± 1.0
± 1.0
± 8.0
± 3.0
± 1.6
± 1.0
± 0.8
± 0.8
± 8.0
± 2.5
± 1.6
± 1.0
± 0.8
± 0.6
± 8.0
± 3.0
± 1.6
± 1.0
± 0.8
± 0.6
± 8.0
± 2.4
± 1.6
± 1.2
± 0.8
± 0.6
Systematic measure-
ment deviation
(inaccuracy) ± [%]
EN ISO
ErgoOne
®
8655-2
± 24
± 8.0
± 4.8
± 4.0
± 2.4
± 2.0
± 16
± 3.0
± 3.2
± 1.0
± 1.6
± 0.8
± 26.7
± 3.0
± 5.3
± 1.0
± 2.7
± 0.6
pipette tips
®
Coincidental measure-
ment deviation
(imprecision; CV) ± [%]
EN ISO
ErgoOne
®
8655-2
± 75
± 12
± 30
± 6.0
± 6.0
± 1.5
± 3.0
± 0.7
± 8.0
± 1.8
± 1.6
± 0.8
± 0.8
± 0.4
± 5,0
± 1.5
± 1,0
± 0.6
± 0,5
± 0.3
± 5,0
± 1.5
± 1,0
± 0.6
± 0,5
± 0.3
± 3.0
± 1.0
± 0.6
± 0.3
± 0.3
± 0.2
± 3.0
± 0.7
± 0.6
± 0.3
± 0.3
± 0.2
± 3.0
± 0.6
± 0.6
± 0.2
± 0.3
± 0.2
± 3.0
± 0.6
± 0.6
± 0.25
± 0.3
± 0.15
Coincidental measure-
ment deviation
(imprecision; CV) ± [%]
EN ISO
ErgoOne
®
8655-2
± 16
± 5.0
± 3.2
± 2.0
± 1.6
± 1.0
± 6.0
± 2.0
± 1.2
± 0.8
± 0.6
± 0.3
± 10
± 1.0
± 2.0
± 0.5
± 1.0
± 0.3