Sample weight

The weight of the sample to be measured must not exceed 15 kg (33 lbs).
1.3.5

Gas cleaning cartridge

The instrument contains a gas cleaning cartridge.
Only a qualified technician shall be permitted to change the gas cleaning cartridge.
Warning! Toxic substances!
The content of the gas cleaning cartridge is toxic.
Compliance with the safety data sheet is mandatory. To request the data
sheet, go to http://MSDS.spectro.com.
1.3.6

Spark stand

The opening of the spark stand plate must be completely covered by the sample
or the respective adapter.
Warning! Optical radiation!
The plasma in the spark stand generates UV radiation. UV radiation may cause
damage to the eyes or skin.
For samples that do not completely cover the spark stand opening, use the
respective sample adapters in such a way that the opening is completely
closed.
Take appropriate protective measures (e.g. wear safety goggles) when the
spark stand opening is not fully covered.
When maintaining the spark stand, compliance with the following information is
mandatory:
Warning! Flammable substances!
Do not vacuum the spark stand. The condensate could ignite an unsuitable
vacuum cleaner.
The exhaust filter of the vacuum cleaner may not be suitable for the pore size
of the condensate.