Fuse Information

Current Input Fuse Replacement

The following table describes the fuse that you should use.
Fuse Part Number Description
EDU34450-36202 Fuse 4 A, 600 V, 6 x 32 mm, Ceramic Fast-Acting Cartridge
If you have determined that the current input fuse is faulty, replace it with one of the same size and rating.
To replace a blown fuse, follow the three steps shown below:
1. The current input fuse can be found in the instrument's front panel. Using a slotted (flat head) screwdriver, turn the
fuse holder clockwise to remove it from the instrument.
2. Remove the blown fuse from the fuse holder and insert the proper replacement fuse into the fuse holder.
Keysight EDU34450A User's Guide
Fuse Type
23