Statistics
The Statistics operation stores the minimum and maximum values, the average, and the number of readings during a
series of measurements.
– This math operation applies to all measurement functions except continuity and diode test.
– The stored statistical data are cleared when statistics are enabled, when the CALculate:FUNCtion command is
sent while the CALCulate:STATe is ON, when the power is turned off, after a Factory Reset (*RST command), after an
Instrument Preset (SYSTem:PRESet command), or after a measurement function change.
Front Panel Operations
Press
[Shift]
> [Null] |
Math
> Function
Statistics
> Statistics
ON
| OFF.
The minimum, maximum, and average readings together with the number of samples taken are shown on the
bottom of the display.
To disable the readings, press Statistics ON | OFF.
Acquire
The EDU34450A can operate in
Continuous
or
Data Log
measurement mode.
Continuous
At power-on, the default measurement mode is Continuous. Continuous readings are taken at the fastest rate
possible for the selected measurement configuration. To set the trigger source for the Continuous mode, follow the
steps below:
Keysight EDU34450A User's Guide