Instructions for Use
For models equipped with turbomolecular pumps of size less than or equal
WARNING
to the Twistorr 305 FS, be sure to block the wheels of the system before
activating the pump so as to avoid accidental movements.
Figure 75 Unlocked wheels
Use inert gas free from dust, particles or humidity (like Nitrogen) for venting the
CAUTION
pump. The pressure at the vent port must be less than 2 bar (above atmospheric
pressure).
Do not use the pumping system in presence of explosive, toxic, fiammable
WARNING
or radioactive gases.