8. Display Mode

When the device is switched on, shortly press the ON/OFF
button 1 to switch to another display mode to select your
desired display mode 9. There are 6 different display
modes. The default setting is mode 1.

9. Using the Lanyard

7
1. Thread the thinner end of the lanyard through the
hanging hole at the rear end of the device.
2. Thread the thicker end of the lanyard through the
threaded end before pulling it tightly.

10. Malfunctions and Actions to take

Descrip-
Symptom/Possible
tion
causes
SpO
or
1. Finger is not
2
pulse rate
inserted
correctly.
do not
2. Patient SpO
display
2
value is too low to
normally.
be measured.
3. There is exces-
sive illumination.
OXY 210
Descrip-
tion
SpO
or
2
pulse rate
is shown
unstable.
The
device
cannot be
powered
on.
Solutions
The
1. Retry inserting the
display
finger.
suddenly
2. & 3. Measure more
switches
times. If you deter-
off.
mine the product is
working correctly,
consult your doctor.
Symptom/Possible
Solutions
causes
1. Finger might not
1. Retry inserting the
be inserted deep
finger.
enough.
2. Sit calmly and retry.
2. Excessive patient
movement.
1. No batteries or
1. Replace the
low battery
batteries.
power.
2. Remove and rein-
2. Batteries are not
stall the batteries.
installed
3. Contact your local
correctly.
Microlife-Customer
3. The device may
Service.
be damaged.
1. The device is
1. Normal.
automatically
2. Replace the
powered off,
batteries.
when no signal
was detected
after 10±2
seconds.
2. The battery
power is too low
to operate.

11. Cleaning and Disinfecting

Use an alcohol swab or cotton tissue moistened with
alcohol (70% Isopropyl) to clean the silicone that touches
the finger inside of the device. Also clean the finger being
tested using alcohol before and after each test. Allow the
device to dry thoroughly before use.
Never use abrasive cleaning agents, thinners or
benzene for cleaning and never immerse the
device in water or other cleaning liquids.

12. Guarantee

This device is covered by a 2 year guarantee from the date
of purchase. During this guarantee period, at our discre-
tion, Microlife will repair or replace the defective product
free of charge.
Opening or altering the device invalidates the guarantee.
The following items are excluded from the guarantee:
 Transport costs and risks of transport.
 Damage caused by incorrect application or non-
compliance with the instructions for use.
 Damage caused by leaking batteries.
 Damage caused by accident or misuse.
 Packaging/storage material and instructions for use.
 Regular checks and maintenance (calibration).
5
EN