Cuff s:Clean the cuff s before examining a patient using the low risk cleaning method below.
Following patient examination, clean and/or disinfect the cuff s by the appropriate method based upon the
level of cross contamination risk. Before fi tting the cuff s to the patient, evaluate the cross-contamination risk
according to the defi nitions in the tables below:
Risk
Defi nitions
Low
Normal use or
low risk situations
including patients
with intact skin and
no known infection.
Medium/
The patient has a
known infection, or
High
skin is not intact.
Refer to the cuff IFU for cleaning guidance and restrictions.
Caution
Do not allow any fl uid to enter the cuff tubing.
Do not use alternative cleaning agents or methods as permanent damage is likely.
Inspect cuff s after cleaning and prior to use.
Cuff Inspection: Cuff s should be regularly inspected. Examine the outer cuff surfaces for material damage,
splitting, fraying etc. Make sure that labelling is clearly legible. Check the cuff tubing and connections for
damage, splits etc. If in any doubt as to the condition, the cuff (s) should be replaced. In any case, cuff s should
be replaced every two years.
Caution
After using chemicals ALWAYS rinse off / remove the chemical with absorbent material, dampened in
clean water and dry with a clean cloth.
8.

Troubleshooting

This table lists some of the common problems found during use, with possible solutions. If the problem cannot
be located after consulting this table, turn off the Doppler and consult a qualifi ed technician.
Before attempting trouble shooting, make sure that the batteries are charged.
SYMPTOM
Doppler will not turn on.
Audio Only
No Audio signal
Poor Signal
No Signal
Screen displays:
Screen displays:
Screen displays:
Procedure
1. Clean with soft cloth and a mild, neutral detergent @ 40°C (104°F)
2. Disinfect using a 70% isopropyl alcohol wipe or chlorine releasing
agent @ 1000ppm available chlorine
3. Wipe with a cloth dampened in clean water.
4. Completely dry with a clean lint-free cloth
Because of the nature of the cuff materials, eff ective cleaning and
disinfection in medium/high risk situations is not practical. Therefore,
we recommend to dispose of according to local procedures
POSSIBLE CAUSE / REMEDY
Replace / re-charge batteries.
Doppler model does not support visual functionality
Incorrect volume setting
Probe / sensor incorrectly positioned, or Insuffi cient Gel
Damaged probe / sensor, or Incorrect probe / sensor
Damaged probe / sensor, or No Probe
Incompatible probe / sensor, or Incorrect Probe / sensor
Incorrect battery fi tted