Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The BM 48 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 48 should
assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
All manuals and user guides at all-guides.com
IEC 60601 test level
± 6 kV contact
± 8 kV air
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
3 A/m
113
Electromagnetic environment –
guidance
Floors should be wood, concrete
or ceramic tile. If floors are co-
vered with synthetic material,
the relative humidity should be
at least 30 %.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment.