Cleaning and Sanitizing,
continued
If filtering beverages, sample residues should be rinsed from the support with hot
water or a mild neutral detergent prior to autoclaving. Do not use oxidizing agents.
The lever mechanism can be disassembled by unscrewing the knurled nut and
pulling the lever out from its socket.
The support and manifold must be autoclaved prior to use. Proper flaming of the
support head is sufficient to prevent cross-contamination from sample to sample.
After every ten autoclavings (or boilings when using the stand-alone support in the
field), deposit one drop of silicone oil on the O-ring after pulling the spring down
a few millimeters. (Do not use a silicone grease, as this could block the venting
system.)
The O-rings should be replaced on a yearly basis. To replace the O-rings of the
integrated two-way valve, disassemble the valve by removing the retaining clip and
pulling on the handle.
Microfil dispenser
This unit is not autoclavable and should not normally require sanitation. If neces-
sary, the funnel magazine may be sanitized with 70% alcohol.
The critical surfaces of the Microfil funnel do not contact any surface of the
dispenser.
The dispenser may be cleaned using a mild detergent. Do not use oxidizing agents.