TROUBLESHOOTING
Type of problem
Control lamps do not illuminate when
the unit is switched on
No liquid flow when the footswitch is
activated
No ultrasonic oscillations when the
footswitch is activated
Decreasing or insufficient ultrasounds
power
Water leakage between handpiece
and handpiece hose
Water leakage
90
FB-128_ed_2007-05_miniEndo.indd 90
Solutions
• Check all the connections of the unit
• Switch on the unit with the "ON/OFF" button
• Check the power supply of the dental office
• Send the unit to an approved SybronEndo repair center
• Turn the liquid flow rate knob
• Check the connections between the handpiece hose and the handpiece
• Check the connections of the unit
• Check the connection of the footswitch
• Check the filter
• Disconnect the handpiece and check with compressed air (max. 3 bar) whether it is clogged.
• If yes, send the handpiece to an approved SybronEndo repair center
• Send the unit to an approved SybronEndo repair center
• Switch on the unit with the "ON/OFF" button
• Check the connections between the handpiece hose and the handpiece
• Check the connections of the unit
• Check the connection of the footswitch
• Check that the instrument is correctly tight
• Increase the power of the ultrasounds to the maximum
• Send the unit to an approved SybronEndo repair center
• Increase the power of the ultrasounds to the maximum
• Check that the instrument is correctly tight
• Check the wear of the instrument and replace it if necessary
• Send the handpiece to an approved SybronEndo repair center
• Check the O-ring of the handpiece
• Send the handpiece to an approved SybronEndo repair center
• Check the water hose connector
• Send the unit to an approved SybronEndo repair center