Recommended Separation Distances - SybronEndo miniEndo Instrucciones De Empleo

Tabla de contenido

Publicidad

Recommended separation distances

The product is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the
product can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the product as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
58
58
FB-128_ed_2007-05_miniEndo.indd 58
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2 √P
0.12 m
0.38 m
1.2 m
3.8 m
12 m
m
800 MHz to 2.5 GHz
d = 1.2 √P
0.12 m
0.38 m
1.2 m
3.8 m
12 m
d = 2.3 √P
0.23 m
0.73 m
2.3 m
7.3 m
23 m
8/14/2015 12:36:24 PM

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido