7. Problems
Problem description
The chiller is not cooling
The chiller is not cooling
The coolant pump is weak
The coolant pump is not
functionning
The display is blinking
Screen shows F1L
Screen shows F1H
EP
Before any servicing operation, please contact :
VWR International BVBA
Researchpark Haasrode 2020
Geldenaaksebaan 464
B-3001 LEUVEN
Belgium
www.vwr.com or your local VWR office
8. Storage and transport
The packaging of the device was carried out with greatest care to make sure that damage by transport
and storage is avoided. Upon receipt, both packaging and device have to be checked for damages.
Should such damages be noticed, please contact the manufacturer.
Should you need to move the chiller, drain the coolant before the transport. The chiller must be dry
before it is moved.
Transport should be done only using the original packaging.
Possible reason
The coolant is dirty
Room temperature is too high
The coolant is dirty
The coolant pump is defective
The coolant temperature is
below or above the allowed
temperature
Temperature sensor issue,
short circuit
Temperature sensor issue,
defective sensor
Reservoir issue
VWR RC-10 Digital Chiller 462-0138 VWR RC-10 Duo Chiller 462-0139
Solution
Maintenance and clean up of
the chiller
Change chiller location
Remplace the coolant
Contact your service personnel
Verify the exact function of the
external device
Compare the temperature
setpoint and the cooling cycle
temperature
Call the customer service
support, replace the sensor
Call the customer service
support, replace the sensor
Turn off the chiller for 5
seconds (if all reservoir
components function properly)
call customer service support
ENG
18