Horizon 6
Horizon 12
Horizon 24
Preset Cycle
Horizon 6 FA
75 & 100 mm tubes
125 mm tubes
R
C
T
EVIEW
YCLE
IME AND
Factory programmed cycles are shown on the rear of the centrifuge, on the Factory Set Cycles label. To review
current settings, follow this procedure:
o The lid must be open to review the selected cycle time and speed.
o Press and hold the START button until you hear a beep.
o Release the START button. The centrifuge will beep and the LED light will flash once for each minute of run time
in the current cycle. 10 beeps / flashes equal 10 minutes of run time. Run time starts when the rotor reaches 90%
of desired speed and stops when the rotor starts decelerating.
o Pressing the START button again will cause the unit to beep and the LED light to flash once for each 100 rpm in
the current cycle. 38 beeps / flashes equal 38 x 100 or 3,800 Revolutions per Minutes (RPM)
o The centrifuge will automatically revert to normal mode at the end.
(1) Chemistry
RPM
Time
G-Force
1800 xg
RPM
Time
G-Force
1800 xg
RPM
Time
G-Force
1800 xg
10 min @ 3900 RPM
1600 xg
1850 xg
S
S
PEED
ETTINGS
Drucker Diagnostics | +1-814-692-7761 |
(2) Coag (PPP)
3,600
3,300
10
15
1500 xg
3500
3200
10
15
1500 xg
3300
3000
10
15
1500 xg
15 min @ 3800 RPM
1500 xg
N/A
CustomerService@DruckerDiagnostics.com
(3) Urine
1,900
5
500 xg
1800
5
500 xg
1700
5
500 xg
5 min @ 2200 RPM
500 xg
600 xg
| 6