Reset to zero
The Geiger counter sensor comes pre-calibrated
however when starting a new experiment with the
"counts" range, a reset option is available.
To zero the sensor:
1. Connect the NUL-247 Geiger counter sensor to a
computer/tablet/smart device following one of the
guides below.
2. Open the NeuLog software.
3. When your sensor has been detected click on the
Geiger sensor module box (on the left side of your
screen)
4. Click on the Extra command button.
5. Click reset to zero your reading.
6. The "counts" value is now set to zero.
7. Alternatively, press on the sensor's push button for
four seconds.