Shenzhen Creative Industry Co. Oxy-One Plus PC-60F Manual De Usuario página 65

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

humidity 10%~95%, atmospheric pressure:
50kPa~107.4kPa.
The Oximeter is calibrated in the factory before
sale, so there is no need to calibrate it during its
life cycle. Any SpO
to validate the accuracy of the Oximeter, they can
only be used as functional testers to verify its
precision. The SpO
manual is supported by the clinical study
conducted by inducing hypoxia on healthy,
non-smoking, light-to-dark skinned subjects in an
independent research laboratory
If it is necessary to verify the precision of the
Oximeter routinely, the user can do the verification
by means of SpO
the local third party test house. Please note that
the specific calibration curve (so called R-curve)
should be selected when use of SpO
for Index 2 series SpO
Biomedical Corporation, please set "Make" to
"DownLoadMake: KRK", then the user can use this
particular R-curve to test the Oximeter. If the SpO
simulator does not contain "KRK" R-curve, please
simulators should not be used
2
accuracy claimed in this
2
simulator, or it can be done by
2
simulator from Fluke
2
15
.
simulator, e.g.
2
2

Publicidad

Tabla de contenido
loading

Productos relacionados para Shenzhen Creative Industry Co. Oxy-One Plus PC-60F

Tabla de contenido