Table 201: Guidance and manufacturer's declaration
T he PERITONE product is intended for use in the electromagnetic
environment specified below. T he customer or the user of the PERITONE
product
should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 202: Guidance and manufacturers declaration
– electromagnetic immunity
T he PERITONE product is intended for use in the electromagnetic environment specified
below. T he customer or the user of the PERITONE product should ensure that it is used in
such an environment, and that precautions regarding that environment are heeded.
Immunity test
IEC 60601
test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
PERITONE Instruction Manual - GB
– electromagnetic emissions
Compliance
Electromagnetic environment –
Group 1
T he PERITONE product uses RF
energy only for its internal function.
T herefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
Class B
T he PERITONE product is suitable
for use in all establishments,
including domestic establishments
Not applicable
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Not applicable
Compliance
level
±6 kV cont act
±8 kV air
3 A/m
28
guidance
Electromagnetic environment–
guidance
Floors should be wood, concrete or
ceramic t ile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic fields should
be at characteristic levels of a typical
locat ion in a t ypical commercial or
hospital environment.