EASY CALIBRATION
Revert to factory default settings
The continually displayed CAL symbol refers to a previously made
adjustment.
What to do
1.
Bring up the CAL
mode
2.
Deactivate CAL
mode
3.
Revert to factory
setting
When the ErgoOne
IMPORTANT:
performed, which means that the volume is changed across
the entire volume range of the pipette by the same amount.
It is recommend that the adjustment be performed at 50%
of the nominal volume.
NOTE:
The instrument is permanently adjusted for watery solutions,
but it can also be set for solutions with varying density,
viscosity and temperature. The ErgoOne
adjusted in every mode, with the exception of the GEL mode.
How to do it
Press and hold the MENU
key (> 3 sec) until CAL ap-
pears. The display reads 'on'.
'CAL' blinks.
Press either the UP or DOWN
key to deactivate the CAL
mode. The display changes
from 'on' to 'off'. 'CAL'
continues to blink.
Press the ENTER key.
The CAL symbol disappears.
The instrument has now
been reverted to factory
default setting.
E -8/-12 is adjusted, a volume offset is
®
Keys to press
1x
>3s
1x
E -8/-12 can be
®
ErgoOne E -8/-12 · User Manual ·