Maintenance
Inspection and
Testing
Packaging for
Sterilization
Sterilization
Temperature
Relative Humidity
Vacuum Set Points
EtO Concentration
Time
•. No.routine.maintenance.is.required.for.the.
sterilization.tray.
•. Inspect.the.tray.for.dents.and.cracks..If.a.problem.
is.observed.or.suspected,.the.tray.should.be.
returned.for.repair.
•. Inspect.all.components.for.cleanliness..If.fluid.
or.tissue.buildup.is.present,.repeat.the.above.
cleaning.and.disinfection.procedures.
•. Double-wrap.the.tray.unless.otherwise.
noted below.
•. Prepare.the.device(s).and.the.sterilization.tray.
as.indicated.in.the."Sterilization.Tray.Setup".
section.below.
•. Sterilize.the.device(s).inside.the.tray.according.
to.the.parameters.indicated.below.
Ethylene oxide (EtO)
Preconditioning
55°C
(131°F)
70%
1.30.psia
60.minutes
Sterilization
55°C
(131°F)
70%
600.mg/L
(100%.EtO)
240.minutes
Aeration
55.±.4°C
(131.±.7°F)
12.hours
EN-7