Warnings
•. These.instructions.are.validated.only.for.sterilization.of.the.tray.and.device(s).
identified.herein..Using.combinations.or.parameters.not.described.in.this.
manual.may.result.in.incomplete.sterilization.
•. These.instructions.do.not.replace.the.cleaning.instructions.provided.with.
individual.devices..Prior.to.sterilization,.clean.all.devices.as.specified.in.their.
respective.user.manuals.
•. Wear.appropriate.protective.equipment.(gloves,.eye.protection,.etc.).when.
reprocessing.any.medical.device.
•. Both.the.device.and.the.tray.must.be.cleaned.prior.to.sterilization,.
or incomplete.sterilization.will.result.
•. The.sterilization.tray,.its.lid,.and.any.internal.components.have.been.designed.
and.validated.for.use.as.a.single.system..Do.not.separate.components.from.
the.system,.for.use.individually.or.in.combination,.or.incomplete.sterilization.
may.result.
•. Inspect.the.tray.and.its.components.for.visible.damage,.such.as.cracking.or.
chipping,.prior.to.use..Do.not.use.the.tray.if.it.is.damaged.
Cautions
•. Before.lifting.the.tray.assembly,.verify.that.the.latches.connecting.the.lid.to.
the.tray.are.secure.
•. The.tray.is.not.designed.for.use.as.a.shipping.container..To.avoid.damage,.
remove.all.devices.and.pack.them.separately.
Limitations on
Reprocessing
•. Proper.processing.has.a.minimal.effect.on.the.
tray..End.of.life.is.normally.determined.by.wear.
and.damage.due.to.use.
•. Do.not.cross-sterilize.the.tray..Using.multiple.
sterilization.methods.may.significantly.reduce.
the.performance.of.the.tray.
•. Do.not.leave.the.tray.in.solutions.longer.than.
necessary..This.may.accelerate.normal.product.
aging.
•. Damage.incurred.by.improper.processing.will.
not.be.covered.by.the.warranty.
EN-3