2.0
Operating the calibrator
2.1
Before use
Warning
•
The calibrator must not be used for any purposes other
than those described in this manual.
•
The calibrator is designed for interior use only and
should not be used in risk-prone areas, where vapour
or gas leaks, etc. may constitute an explosives hazard.
•
The connections used to test thermostats (Fig. 1 pos. 8
and 9) must NEVER be connected to voltages
exceeding 50V. The thermostats must not be connected
to any other source of voltage during the test.
•
NEVER use heat transfer fluids such as silicone, oil,
paste, etc.
•
The calibrator must be kept clear within an area of 20
cm on all sides and 1 metre above the calibrator.
Caution – Hot surface
This symbol is engraved in the grid plate.
•
Do not touch the grid plate, the well or the insertion
tube as the calibrator is heating up – they may be very
hot.
•
Do not touch the handle of the calibrator during use – it
may be very hot.