Sterilisation

The handles must be packaged in compliance with EN 868-5.
The handles can be sterilised with standard cycles 121°/134° C up to two hundred (200) cycles
or however up to loss of the mechanical performance.
The parameters of the sterilisation cycle are as follows:
Ciclo EN
13060
B
B

8. PERIODIC CHECKS

Operation
Check the absence
of range between the
arm joints
Check data plate is
legible
Check casing
Twice-yearly
is intact
Check electrical safe-
Twice-yearly
ty EN 62353
1. Rigidity
2. Dispersion
Light parameter
Twice-yearly
checks.
Service Engineer: competent person in electro-mechanical equipment maintenance.

9. ACOUSTIC ALARM SIGNALS

9.1

Acoustic Signals

OpL** = Beep 30 seconds
OTP* = Beep 30 seconds
* OTP: LED overtemperature protection.
** OpL: Disconnected led load
Temperature
121°C
134°C
Applicability
Frequecy
LD
TE
Annual
x
N/A
Annual
x
x
x
x
x
x
Pressure
Minimum
Holding Time
207 KPa
15 min
308 KPa
3 min
Procedure
Check the gap between the
joints 5 and the arm has not
changed since first use
Measure dielectric rigidity and
dispersion of the casing.
Limits defined in IEC 60601-1
With a spectroradiometer,
measure the values of:
• Maximum lighting: >35000
lux
• CRI decay: <20%.
• Underlying value of the blue
light on the spectrum emitted
measured in: <100 W/m2
Enabling
User
User
Service
Engineer
Service
Engineer
Service
Engineer
63