• The face masks and the air hose must not be placed in hot water!
• The device must not come into contact with water or heat!
The face mask should be rinsed out with water after use. The device and the
air hose should be wiped out once a day with a clean, moist cloth – the
device must not be plugged into the electricity supply during this
process!
For doctors' surgery or hospital use:
The nebuliser and mouthpiece can be placed in an autoclave at maximum
121 °C for 30 minutes. Face masks, air hose and device are not suitable
for hot water or the autoclave!
4. Maintenance, care and service
You can procure all necessary spare parts from your dealer or pharmacist, or
please contact Microlife directly for details of your local Microlife supplier.
You will find detailed user information on our products and services at
www.microlife.com
We recommend replacing the nebuliser, mouthpiece and masks after 3 months
use.
The filter should be continually
checked for cleanliness and replaced if
dirty, or after a maximum of 3 months
use. Spare filters are provided with the
device. To replace the filter, open the
compartment on the device marked
«Open» and exchange the filter as
shown.