INSTRUCTIONS FOR USE
c. To prevent patient and/or user injury and/or equipment damage, examine the
device and surgical-table side rails for potential damage or wear prior to use. Do
not use the device if damage is visible, if parts are missing or if it does not function
as expected.
d. To prevent patient and/or user injury and/or equipment damage, verify the
device attaching clamps completely touch the table-side rails and are firmly in
place. Test the locking mechanism to ensure no movement when elevated or
pushed.
e. To reduce the chances of patient tissue damage, care should be taken to ensure
proper orientation and location of supports.
f. To reduce the chances of skin maceration and shear, it is recommended that a
disposable cover be used between the patient and the pads.
g. To avoid injury to patient, ensure Wing Sets are properly latched and locked prior
to loading a patient.
CAUTION:
a. Do not exceed safe working load shown in the product specification table.
b. To avoid injury to patient, ensure Wing Sets are properly latched and locked prior
to loading a patient.
Product Specifications:
Mechanical Specifications
Product Dimensions
Material
Safe Working Load on the device
Overall Weight of Complete Device
Storage Specifications
Storage temperature
Storage Relative humidity range
Operating temperature
Operating Relative humidity range
Electrical Specifications
Not Applicable.
Software Specifications
Not Applicable.
Document Number: 80028214
Version: A
5" X18 ½ "X 3 7/8"(12.7cm x 47cm x 10cm)
Polyethylene, Urethane, Acetal, Nylon, High
Density Foam, Low Density Foam, Acrylonitrile-
Butadiene-Styrene, Garolite
500 lbs. (226 kg)
8 lbs. (3.6 kg)
-29ᵒ C to +60ᵒ C
15% to 85%
This device is intended to be used in a
controlled Operating Room environment.
Not Applicable.
Not Applicable.
Page 15
Description
Description
Description
Description
Issue Date: 25 MAR 2020
Ref Blank Template: 80025118 Ver. E