OPERATING INSTRUCTIONS
M2000 is a universal mixer for amalgam, glass ionomer cements and other pre-dosed dental materials in capsules and syringes.
The apparatus has been developed solely for use in dentistry. Liability cannot be accepted for damage resulting from misuse or failure to observe the
Operating Instructions. The user is responsible for testing the M2000 for its use and suitability for the intended purposes. The intended use also
includes the observation of the notes and regulations in these Operating Instructions. The M2000 has been shipped from the manufacturer in a safe
and technically sound condition. The apparatus is built according to: Directive 93/42/EEC and subsequent amendments contained in according with
the Directive Rule 89/336/CEE, 92/31/CEE and 98/68/CEE and the following harmonized standards have been applied: EN 61010-1 - EN61010-2-81-
EN61326-1 - EN61000-3-2 – EN6100-3-3 - ISO 7488 :1991
CAUTION !
Please read these instructions carefully before you set up and start the device. As with all technical equipment, this unit will function properly and be
safe to operate only if both the general safety regulations and the specific safety instructions laid out in these Operating Instructions are followed.
1. The unit must be operate accordance with the following instructions. We accept no liability for any damage arising from use of the unit for any
other purpose.
2. Prior to setting-up the device, ensure that the voltage stated on the rating plate is identical to the available main voltage. Use of the device at
different voltages may lead to the destruction of the unit.
3. The unit must be operated with the cover intact and closed, otherwise, there is a risk of injury.
4. The cover must not be opened until the capsule holder has come to a complete stop.
5. Only the capsule to be mixed but no other objects may be present in the space below the cover prior to initiating the mixing cycle.
6. Placement or operation of the unit on non-level surfaces is not permissible.
7. For operation of the unit, connect the main plug to an adequate electrical socket with grounded lead. If an extension cable is used, ensure that the
grounded lead is not interrupted.
8. In order to prevent electrical shock hazards do not insert any objects into the unit with the exception of replacement parts properly installed in
accordance with the present instructions.
9. The device should be opened and repairs be carried out only by qualified personnel.
10. Only adequately labelled original NORDISKA DENTAL replacement parts may be used in the repair and maintenance of this unit. Any damage
resulting from the use of third-party replacement parts voids the liability of NORDISKA DENTAL.
11. Do not expose the device to solvents, flammable liquids, and strong heat sources as these may damage the plastic housing.
The corresponding instructions for use of the material suppliers have to be observed.
If it has to be assumed that safe operation is no longer possible, the power must be disconnected to avoid accidental operation. This may be the
case, for example, if the apparatus is visibly damaged or no longer works correctly. A complete disconnection from the power supply is only ensured
when the power cord is disconnected from the power source.
For safety during transit, the capsule mixer has been fitted with a transit screw prior to shipping. It is essential to remove this screw prior
to using the capsule mixer.
a) Place the unit on its side on a soft surface
b) Using the supplied 3mm allen key, turn the transit screw anti-clockwise until it is loosened. Remove screws.
c) Place the capsule mixer correct side up on a flat surface in an easily accessible
Packaging part list: Electronic Control Console -Power Cord – Allen Key – DFU (Fig. D)
1. Connect the power cord with the socket at the rear of the apparatus and subsequently with the mains socket. (Fig. B)
2. Switch on the apparatus at the rear. (Fig. B)
3. Raise the safety cover and place the mixing capsule in the mixing fork (Fig. C)
- Insert the capsule in the right cap of plastic fork (Fig. C-1)
- Open the fork moving the left part (max 8 mm) (Fig. C-2)
- Turn the capsule inside the plastic fork (Fig. C-3)
- Check the capsule is clamp correctly.
Manual Working
1. Plan the wished time with the bottoms "+" and "-"
2. Push the bottom "S" (Start). When the operation is finished, it will remain planed.
If the safety cover is opened during the mixing process, the circuit is broken and the motor stops automatically. The mixing process might be
finished by opening the safety cover or by pressing the button "S" again. When you turn off the main switch, the amalgamator remains planed on
the last operation it executed. M2000 fits capsules between 30 and 36 mm of length and with 14/18 mm of diameter.
DFU – M2000 Rev. 0 - 09/2016
INTENDED USE
DENTAL MATERIALS USED
ASSUMPTION OF IMPAIRED SAFETY
REMOVAL OF TRANSIT SCREWS
BEFORE USING REMOVE TRANSIT SCREW FROM UNDERSIDE OF UNIT. (Fig A)
NORDISKA DENTAL AB
Page 2 di 12
SAFETY
location.
USE