neral switch (rif. 28) and regulate the aspiration power acting on the power regulator (rif. the 29) min 400
W - max 1000W.
ATTENTION: the AIRBAG filter (rif. 14) has a very important function for the perfect operating of the appa-
ratus and for your health. Check it periodically and replace it before it is completely full. Do not use the
same filter more than one time, and do not discharge it. Non-observance will void your warranty.
Once exausted the filter close it with a proper adhesive tape and waste appropriately in compliance with
your country rules.
IMPORTANT INFORMATION
The Manufacturing Company declines any responsibility for any aparatus installed, utilized, modified, re-
paired and/or altered by non competent and unauthorized personnel.
The aparatus has been designed and built only for the uses contemplated, consequently the Manufactu-
ring Company refuses any responsibility for damages to persons, animals, things, and the machine itself
caused by improper use of this aparatus.
The improper use of this machine is dangerous!!
Subject the aparatus to a complete annual revision by the Manufacturer.
To clarify any doubt and/or technical explanation please contact our Technical Department.

MAINTENANCE

Substitution of the used filter of powders:
Open the lateral panel (rif. 33) of the suction unit removing the two screws (rif. 32), control the bag of used pow-
ders (AIRBAG - pack 4 pcs. - rif. 14) every 4/5 days and replace it before completely full (max to approximately
2/3 of its capacity). Place a new filter bag making a light pressure upwards untill it is perfectly fitted in the ple-
xiglass connection (rif. 5). Do not re-use and do not empty out the used filters. When the filter is exausted close
it with a proper adhesive tape and waste appropiately in compliance with your country rules.
Substitution of the motor protection filter:
Remove the metal central panel screw (rif. 30), rotating the metal holder disk (rif. 31); pull and extract upwards
the filter. Insert a new filter (rif. 12 - cod. 3106012) and close the all making operations to the contrary. Check
every two weeks the motor protection filter and clean it by blowing compressed air in the same way of passage
of the sucked air. The filter must be replaced with a new one at least every 4 months. This preserves the motor
and the perfect operating of the apparatus.

GUIDE TO COMMON QUESTIONS

The apparatus is not turning on:
1
Control that the power cord plug (rif. 1) is properly connected to the electrical outlet.
2
Check that fuses (rif. 24) are not damaged.
The suction power is low:
1
Replace the recovery dust filter (rif. 14), and/or replace the motor protection filter (rif. 12).
2
Check that the motor is correctly placed in the center of its site (rif. 17a23).
Carried out these controls, in case the apparatus still presents disadvantages, contact the customers' technical ser-
vice.

SPECIFICATIONS

Fuses: 2 fuses 5x20 mm. 6,3 Ampere
Power max 1000 Watt.
Operating voltage V/Hz 230/50-60 (see nameplate on the apparatus)
Weight Kg. 11
DIMENSIONS GS-EMC
Height: cm. 40
Width: cm. 33
Depth: cm. 25
DIMENSIONS TWIN
Airbox