ADAPTER INSTALLATION INSTRUCTIONS
*
Important: If Den-Tal-Ez unit is manufactured after January 1997, syringe will require an A-1 Conversion Kit.
1
2
3
USER INSTRUCTIONS
1. Remove existing
thumb nut and
syringe tip.
2. Verify new adapter
n
has o-rings
and
o
in position as shown.
3. Hand-thread adapter
into syringe. Use hex
key provided to tighten.
1
1. With thumb nut loosened, insert new Sani-Tip disposable tip and
push into adapter as far as possible. While still pushing inward, twist
tip 1/2 turn clockwise to create a watertight seal over internal tapered
water spindle.
2
2. Hand tighten thumb nut. To redirect tip, push in on tip while
rotating clockwise.
TEST PRIOR TO EACH USE: Express water, then express air only.
If air contains moisture, or tip dislodges, refer to Troubleshooting
section.
3. To replace tip, loosen thumb nut. Remove and discard used tip.
Repeat steps 1 and 2.
p
4
4. Verify o-ring
m
place inside large
opening of white
plastic cone
l
5
5. With o-ring in place,
firmly snap large end
of white plastic cone
over adapter rim.
6
6. Hand-thread thumb
nut onto adapter. Do
not tighten until new
tip has been inserted.
Installation is complete.
is in
.
3