Determining/Setting the cell constants
C
A L
1/
cm
0 7
4
3
Tref25
Cal
E
L
C L
6
8 0
S/
cm
1/
cm
0 1
6
4
Tref25
ARng
nLF
C L
E
L
9
2 5
S/
cm
1/
cm
0 1
0
0
Tref25
ARng
nLF
8
All manuals and user guides at all-guides.com
! Press <RUN/ENTER>. The AutoRead measurement begins.
! If the measured value is stable, the instrument displays the determined
cell constants and the calibration evaluation. The measuring instrument
automatically stores the cell constants.
! Switch to the measuring mode with <M>.
Procedure for manually setting the cell constant:
! Press the <CAL> key repeatedly until the
! Then press the <RUN/ENTER> key.
! Press the <CAL> key until a measured value and a value for the cell con-
stant appears in the required setting range.
! Set the value of the cell constant with <L> <M>.
! Switch to the measuring mode with <M>.
Procedure for selecting fixed cell constants 0.010 cm
! Press the <CAL> key repeatedly until the
! Then press the <RUN/ENTER> key.
! Press the <CAL> key until a measured value and the cell constant
-1
0.010 cm
appear.
! Switch to the measuring mode with <M>.
BYK LC 3
CELL display appears.
-1
:
CELL display appears.