Zero point adjustment (OFFSET)
Zero point adjustment (OFFSET)
The gage should be verified before each series of measurements by using a gage block. 
If there is a deviation from the specified measurement the device should be re‐adjusted 
via SETUP‐Menu OFFSET. 
 
Always the total selected OFFSET value will be shown! 
The value by which the gage must be adjusted has to be added or 
subtracted. 
In order to increase the accuracy of the measurement, the gage should 
be calibrated in the same position in which later the measurement will 
be effected. 
 
The instrument must be switched on.
program
 1x long
 
 1
x short 
 1x long
 
 
short 
 1x
 short 
 
short 
 
short 
 1x 
short 
Repeat step 6‐8 until the OFFSET value is totally entered.  
10 
 1x long
 
 
11 
 
 
The SETUP – Menu can be activated in each 
 
Activating the SETUP – Menu  
„OFFSET" will appear in the display 
Activating the OFFSET – Menu 
Selecting the sign plus (+) / minus (‐) 
Press key several times 
Selecting the first digit to be settled 
Increasing the first digit or..... 
 
.....Reducing the first digit 
 
Selecting the next digit 
For scale interval 0,005  3 digits are editable! 
For confirming and saving 
Afterwards you will return to the SETUP ‐ 
Menu. 