Problem/
Possible cause/remedy
question
Why should
There are two reasons for this:
the atomiser
1. To ensure a therapeutically effec-
be replaced
tive particle spectrum, the nozzle
regularly?
hole must not exceed a specific
diameter. Due to mechanical and
thermal stresses, the plastic is
subject to a certain degree of
wear. The atomiser insert [11] is
particularly sensitive. This can also
alter the composition of the aero-
sol droplets, which directly affects
the effectiveness of the treatment.
2. Regularly changing the atomiser
is also recommended for hygiene
reasons.
Should each
Yes, this is absolutely essential for
person have
hygiene reasons.
their own
atomiser?
10. Technical specifications
Model
AERO-N
Type
GCE834
Dimensions
(W x H x D)
110 x 62 x 47 mm
Weight
222 g
Operating
pressure
Approx. 0.25 to 0.5 bar
Atomiser filling
Min. 2 ml
volume
Max. 6 ml
Medication
flow rate
Approx. 0.25 ml/min.
Sound pressure
Max. 45 dBA
(acc. to DIN EN 13544-1 section 26)
Mains connection Input: 100 – 240 V~; 50 – 60 Hz; 0.5 A
Output: 5 V
Expected
400 h
service life
Operating
conditions
Storage and
transportation
conditions
Aerosol properties 1) Aerosol delivery: 0.15 ml
Battery:
Capacity
Nominal voltage
Type designation
The serial number is located on the device or in the bat-
tery compartment.
Subject to technical changes.
Particle size diagram
Measurements were performed using a sodium fluoride
solution with a "Next Generation Impactor" (NGI).
The diagram may therefore not be applicable to sus-
pensions or highly viscous medications. You can obtain
more detailed information from the manufacturer of your
medication.
; 2 A
23
Temperature: +10 °C to +40 °C
Relative humidity: 10% to 95%
Ambient pressure: 700 to 1060 hPa
Temperature: -20 °C to +60 °C
Relative humidity: 10% to 95%
Ambient pressure: 700 to 1060 hPa
2) Aerosol delivery rate: 0.03 ml/min
3) Particle size (MMAD): 4.12 µm
2570 mAh
3.7 V
Li-Ion