Prodeco AERO-N Instrucciones De Uso página 50

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Table 3
Guidance and manufacturer's declaration – electromagnetic immunity
The GCE834 is intended for use in the electromagnetic environment specified below.
The customer or the user of the GCE834 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Conduced RF
150kHz to
IEC61000-4-6
80MHz:
3V
rms
6 V
(in
rms
ISM
and amateur
radio bands)
80% Am at
1kHz
Radiated RF
10V/m
IEC61000-
80% Am at
4-3
1kHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the
GCE834 is used exceeds the applicable RF compliance level above, the GCE834 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the GCE834.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 4
Recommended separation distances between portable and mobile RF communications equipment and the GCE834
The GCE834 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user
of the GCE834 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF commu-
nications equipment (transmitters) and the GCE834 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment should be used no closer to any part of
the GCE834 including cables, than the recommended separation distance calculated from the
equation applicable to the frequency the transmitter.
150kHz to
Recommended separation distance:
80MHz:
3Vrms
d = 0.35
6 Vrms (in
ISM
d = 1.2
80 MHz to 800 MHz
and amateur
radio bands)
d = 1.2
800 MHz to 2.7 GHz
80% Am at
1kHz
d = 2,3
Where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in metres (m).
10V/m
80% Am at
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,
1kHz
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
(out ISM and amateur
radio bands)
d = 1.2
0.12
0.38
1.2
3.8
12
(m)
150 kHz to 80 MHz
(out ISM and amateur
radio bands)
d = 0.6
0.06
0.19
0.6
1.9
6
50
b
80 MHz to 800 MHz
d = 1.2
0.12
0.38
1.2
3.8
12
a
800 MHz to 2.7 GHz
d = 2.3
0.23
0.73
2.3
7.3
23

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido