INSTRUCTION MANUAL
Time/Date
PID/Autotuning
Comunication
Reject change
Validate change
J
.P. SELECTA s.a.u.
CODE 80308 REV 0
7.2 Setting screen
Icons meaning:
7.2.1 Set time and date
Press on the time/date icon. Press successively on the box to select the day,
month, year, minute and second adjustment.
Adjust by pressing on the icons and decrease.
7.2.2 Set parameters of temperature control and
«Autotuning»
For an optimal use of the oven, P, I, D parameters of temperature control
should be the appropriate value.
With a value of the appropriate P, I, D parameters, the oven will reach the
temperature in the shortest possible time and without exceeding the set
temperature. However, those values can be quite different at high and low
temperatures.
The incubator comes with the following values: P=10; I=100; D=25 which
are appropriate for most jobs.
These values can be modifi ed, but it's advisable to perform an «Autotuning»
when you need a better accuracy.
The «Autotuning» automatically adjusts the optimal parameters for a given
temperature and load.
Press on the PID/Autotuning icon. Press successively on the box to select
the P, I, D values adjustment and modify the value by pressing on the In-
crease and Decrease icons.
Attention: It is not recommended to change these parameters. Use the
«Autotuning» option to automatically adjust these values.
Press on the PID/Autotuning icon.
Select the «Autotuning» temperature with:
(for example, the temperature most used)
Start «Autotuning» by pressing on the icon
If for some reason you want to stop, press on:
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Feb/2015
(It can be modifi ed without notice)
ºC/ºF
Page: 16
Alarms and temp. offset
Temperature units
Language (Spanish, English,
French)
Up/Down