Acculan Battery GA 616, GA 626, GA 646
Dear Customer,
With the Acculan system you have also obtained a very
powerful battery with integrated electronic. Due to its
construction and the computerized charging unit the
battery has a very long life expectancy if it is treated
properly.
Therefore we would ask you to observe the following
instructions:
CAUTION!
Ø
The temperature of the battery should not exceed
65 °C.
If the battery is heated over this temperature, the
battery loses capacity, and it will even be destroyed
if the heating lasts longer.
Ø
Avoid a short-circuit.
If the poles of the battery (either top side or down
side) are connected by metal or liquid, high current
will flow (comparable with a car battery). The bat-
tery may be heated up extremely and be destroyed
and it may even damage surrounding objects
(danger of fire).
Therefore we recommend to treat the battery as
follows:
Ø
Store the battery at room temperature (about
20 degrees celsius).
Ø
Do not sterilize or desinfect the battery by any
means.
As the battery is completely sealed inside the hous-
ing of the machine the contamination of the OR-
field by the battery is not possible.
Ø
Always keep the battery dry.
Ø
Never immerse the battery in fluids.
Ø
Never place the battery on wet surfaces.
Ø
Always make sure that the sockets of the charging
unit are dry inside before you put the battery in.
Ø
Do not place or run the charging unit near to places
where you handle fluids (instrument cleaning, infu-
sions).
If you follow these instructions, we assure you will
have many successful operations with your Acculan
system.
If you have any additional questions, please call your
local AESCULAP sales representative.
Keep dry
Disposal
The NiCd cells of rechargeable batteries GA 626 and
GA 646 contain heavy metals and may not be disposed
of with conventional refuse.
Ø
The NiCd cells of rechargeable batteries GA 626 and
GA 646 should be recycled in an environmentally
safe manner.
The NiMH cells of rechargeable battery GA 616 contain
no heavy metals.
Ø
The NiMH cells of rechargeable battery GA 616 may
be disposed of with conventional refuse.
Application
For information regarding the use of the rechargeable
batteries, please consult the appropriate instructions
for use.