The instrument will enter measure mode in the last parameter used. To switch off the instrument, press
the key
in measure mode.
Replacement of batteries
The instrument works with 3 AA 1.5V batteries. To proceed with the replacement:
1.
Turn off the device.
2.
Turn the instrument over with the display facing down and place it on a stable surface.
It is advisable to put a cloth to avoid any scratching on display.
3.
Using the screwdriver supplied, completely unscrew the screw close to the battery symbol.
4.
Remove the battery stopper cap with the help of the lanyard.
5.
Remove the 3 exhausted batteries (one in the left compartment and two in the right compartment)
and insert the new ones. Pay attention to the correct polarity. Follow the diagram above the battery
symbol in the rear compartment of the instrument.
6.
Reinsert the battery stopper cap; always hold it with two fingers, insert the screw and tighten.
Instrument transportation
The instrument is always supplied with the appropriate carrying case. Use the original case only, to transport
the instrument. If you need to buy it again, contact your local distributor.
The interior of the case is shaped to be able to house the instrument and the sensors still connected.
Key functions
Button
Pression Function
Press to turn the device on or off.
Short
In measure mode, press to scroll through the different parameters:
Short
Short
In measure mode, press to enter the setup.
Short
In the setup menus, press to select the desired program and / or value.
During calibration, press to confirm the value.
In the setup and subset menus press to scroll.
In the setup submenus, press to change the value.
In memory recall mode, press to scroll through the saved values.
In MTC and custom calibration mode, press to change the value.
Short
the recording (Automatic Data Logger).
In measure mode, keep one of the two keys pressed to change the temperature in MTC
Long-
mode (manual compensation, without probe). When the value starts to flash, the user
press (3s)
can change the temperature value by entering the correct one and confirming with
UM Serie 70Vio EN rev. 1 04.06.2020
pH 70 Vio : pH mV ORP
COND 70 Vio: Cond TDS Sal
PC 70 Vio: pH mV ORP Cond TDS Sal
In calibration, setup and memory recall mode, press to return to measure mode.
In measure mode, press to start the calibration.
: In measure mode, press to save the data (manual Data Logger) or start and end
: In measure mode, press to recall the saved data.
.