 Manual COND calibration
Example: calibration at 5.00 µS / cm with sensor with Cell Constant 0.1
Access the Setup menu for COND SETTINGS and select in the P3.1  0.1 and in P3.2  Custom,
return to measure mode and go to COND mode
Press the key
to enter the calibration mode.
Rinse the cell with distilled water and dab gently with paper towel.
Apply few ml of standard solution and dip the sensor in the conductivity standard 5.00 µS/cm.
The string "WAIT FOR STABILITY" and the flowing red bands indicate that the measurement is not stable
yet.
Wait for the Conductivity value on the display to stabilize; when the icon
and
to adjust the value by entering that of the standard solution (e.g.: 5.00 µS/cm), as
indicated by the string "ADJUST THE VALUE" and by the icon
When the icon
appears again, confirm the calibration point by pressing the key
The actual measured value flashes on the display and the updated cell constant is then shown.
The icon
appears, which indicates that the instrument is calibrated in the low conductivity range.
Automatically, the device returns to measure mode.
For each cell constant (P3.1) the instrument stores the calibration in order to allow the user, who uses
multiple sensors with different constants not to be forced to recalibrate each time.
The instrument recalls the last calibration with the respect to the selected parameters P3.1 (cell
constant) and P3.2 (type of calibration solutions).
Note: If you are not aware of the exact compensation coefficient, to obtain an accurate
calibration and measurement set in P3.4
exactly to the reference temperature. Another method of working without temperature
compensation is to use the appropriate
thermal tables shown on the most Conductivity solutions.
Important: Always rinse the cell with distilled water before calibration and when switching
from one standard solution to another to avoid contamination.
Replace standard solutions frequently, especially low conductivity ones.
Contaminated or expired solutions can affect the accuracy and precision of the measurement.
 Errors during calibration
NOT STABLE: The button
appear, to confirm the first point.
WRONG BUFFER: The buffer you are using is polluted or not part of the recognized families.
CALIBRATION TOO LONG: The calibration has exceeded the time limit, only the points calibrated up to
that moment will be stored.
 Performing Conductivity measurement
Access the Setup menu for Conductivity to check the calibration, and if necessary, update the reading
parameters; press key
Press
to scroll through the different screens of parameters until activating the Conductivity
parameter indicated by the icon
UM Serie 70Vio EN rev. 1 04.06.2020
.
0.00 %/°C and then work by bringing the solutions
has been pressed with unstable signal. Wait for the icon
to return to measure mode.
.
27
appears, use the keys
.
EN
.
to

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pc 70 vio

Tabla de contenido