OPOX 11-3
Measuring ranges,
accuracy
Storage conditions
ba77078d03
05/2021
Variable
Measuring range (max. resolu-
tion)
Measuring
0.00 ... 20.00 mg/l O
ranges at
(± 0.01 mg/l)
20 °C (68 °F)
20.00 ... 50.00 mg/l O
(± 0.01 mg/l)
0.0 ... 200.0 % O
(± 0.1 %)
200.0 ... 500.0 % O
(± 1 %)
0.0 ... 400.0 mbar O
pressure (± 0.1 mbar)
400.0 ... 1000.0 mbar O
pressure (± 0.1 mbar)
Temperature
-5 ... 70 °C (± 0.1 °C)
Recommended storing
method
Storage temperature
Precision of temperature
measurement
Working life of the sensor
cap
2
2
saturation
2
saturation
2
partial
2
partial
2
In the calibration and storage beaker
0 ... 50 °C (32 ... 122 °F)
± 0.2 K
Min. 1 year with authorized use
EN
English
Accuracy (± 1 Digit)
± 0.10 mg/l or
± 1 % of the measured
value)
± 10% of measured
value
± 1.0 % O
saturation
2
or
± 1 % of the measured
value (depending on
which value is
greater)
± 10% of measured
value
± 2.0 mbar or
± 1 % of the measured
value
(depending on which
value is greater)
± 10% of measured
value
± 0,2 °C