4. Instructions For Use Con't.
2. Hollow Wrapped Cycles
The Statim 2000S and 5000S
feature two WRAPPED 134˚C type S
sterilization cycles.
To select one of these S Cycles:
Press the WRAPPED button to
scroll through the available cycles.
HOLLOW WRAPPED (S) 134˚C / 3.5 min
HOLLOW WRAPPED (S) 134˚C / 18 min
Once the desired cycle has been selected, press the START button.
The unit remembers the last cycle run and then displays this first when the unit
is powered on.
A Process Challenge Device (PCD) is available for validation of HOLLOW WRAPPED (S)
134˚C/3.5 min cycle.
3. Rubber and Plastics Cycle
The Statim 2000S and 5000S feature
121˚C type S sterilization cycles.
To select one of these S Cycles:
Press the RUBBER / PLASTIC button
to scroll through the available cycles.
RUBBER / PLASTIC (S) 121˚C / 15 min
RUBBER / PLASTIC (S) 121˚C / 30 min
A
B
A
100˚C / 101 kPa
B
6 purges (Conditioning)
peak at 133˚C / 295 kPa
vent to 115˚C / 169 kPa
C
Pressurizing
A
B
C
A
100˚C / 101 kPa
B
3 purges (Conditioning)
peak at 120˚C / 199 kPa
vent to 110˚C / 143 kPa
C
Pressurizing
17
C
D
E
F
3.5
min /
18 min
D
Sterilizing
134˚C / 304 kPa -
138˚C / 341 kPa
E
Venting
F
Air Drying
two
D
E
F
15
30
min /
min
D
Sterilizing
121˚C / 205 kPa -
124˚C / 225 kPa
E
Venting
F
Air Drying