4. Fix the adapter.
(Fig. 4)
5. Mount the dispenser-
Screw it to the reservoir by applying gentle hand torque applied to
the threaded platform base only. In case of removal, apply the same
technique to the base, in the opposite direction.
(Fig. 5)
6. Ready to Use
Do not operate the piston until the unit is safely and fully
mounted on the reservoir bottle.
Intended Use
VWR Bottle Top Dispenser is a general purpose laboratory instrument intended for use in
laboratories for dispensing reagents and chemicals which are compatible with the
instrument.
For research use only. Not intended for any animal or human therapeutic or diagnostic use.
Symbols and Conventions
CAUTION This symbol indicates a potential risk and alerts you to
proceed with caution.
Product Specifications
Volume
Range (ml)
0.25-2.5
0.5-5
1-10
2.5-30
5-60
10-100
6. vwr.com I Instruction manual VWR
Specifications as per ISO 8655
Increment
(ml)
±%
0.05
0.5
0.1
0.5
0.2
0.5
0.5
0.5
10
0.5
20
0.5
Bottle Top Dispenser with Recirculation Valve
®
Accuracy
ml
0.0125
0.025
0.050
0.150
0.300
0.500
Fig. 4
Fig. 5
CV
±%
ml
0.2
0.005
0.2
0.010
0.1
0.010
0.1
0.030
0.1
0.060
0.1
0.100