Sample preparation
22
Sample preparation
Mix the standard solutions and test samples 1:1 with TISAB sample condi-
tioning solution. This sample conditioning solution creates optimum condi-
tions for measuring. It provides a constant ionic strength and similar diffusion
potentials at the reference electrode in standard solution and test sample.
If you would like to have more detailed information concerning sample prep-
aration and measuring procedures, WTW provides a large number of appli-
cation reports for various applications.
Response times
The response time depends on the concentration range. It is
several seconds at high concentrations,
several minutes near the detection limit.
The measured value is stable if the value does not change by more than
0.1 mV within 30 seconds.
Interferences
pH values < 5 or > 7
F 500 / F 800
ba15358e03