ZEISS
Ambient conditions
Transport (in packaging):
Permissible ambient temperature .............................................................................. -40 °C to +70 °C
Storage:
Permissible ambient temperature ............................................................................. +10 °C to +40 °C
Permissible relative air humidity (no condensation) ................................................ max. 75 % at 35 °C
Operation:
Permissible ambient temperature ............................................................................. +10 °C to +40 °C
Permissible relative air humidity (no condensation) ................................................ max. 75 % at 35 °C
Altitude of operating site ................................................................................................ max. 2000 m
Atmospheric pressure ......................................................................................... 800 hPa to 1060 hPa
Degree of pollution .......................................................................................................................... 2
Operating data
Operating area ............................................................................................................... Enclosed rooms
Protection class ..................................................................................................................................... I
Ingress protection rating .................................................................................................................. IP 20
Electrical safety .......................................................... in compliance with DIN EN 61010-1 (IEC 61010-1)
.............................................................................................
Overvoltage category ............................................................................................................................ II
Radio interference suppression ........................................................ in compliance with EN 55011 Class B
Noise immunity ................................................................................ in accordance with DIN EN 61326-1
Line voltage of controller module ..................................................................... 100 to 240 VAC (±10 %)
Line frequency ...................................................................................................................... 50 to 60 Hz
Power consumption of controller module ...................................................................................... 70 VA
Colibri 5/7 overall LED classification ....................... LED risk group 3 according to DIN EN 62471:2009
Solid-state light source Colibri 5,
type RGB-UV ............................................................ LED risk group 3 according to DIN EN 62471:2009
Solid-state light source Colibri 7,
type RGB-UV ............................................................ LED risk group 3 according to DIN EN 62471:2009
Solid-state light source Colibri 7,
type RYB-UV ............................................................. LED risk group 3 according to DIN EN 62471:2009
Solid-state light source Colibri 7,
type R[G/Y]B-UV ....................................................... LED risk group 3 according to DIN EN 62471:2009
Solid-state light source Colibri 7,
type R[G/Y]CBV-UV ................................................... LED risk group 3 according to DIN EN 62471:2009
Solid-state light source Colibri 7,
type FR-R[G/Y]BV-UV ................................................ LED risk group 3 according to DIN EN 62471:2009
8
INSTALLATION CONDITIONS
The installation conditions of the microscope used are valid. Otherwise, the ambient conditions from the
chapter Technical data must be observed.
6
INSTALLATION CONDITIONS
423052-6044-008
and conforming to CSA and UL regulations
Colibri 5/7
09/2017