Inserting The New Saw Blade - HEBU medical TITANIUM II Instrucciones Para El Uso Y El Servicio

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN

5.2 Inserting the new saw blade

1. Check that the female thread and saw blade support of the drive shaft are clean.
2. Position the new saw blade.
It makes no difference which side of the saw blade is facing inwards or outwards.
Important!
The notches in the saw blade must be positioned precisely flush on the spigots of the
drive shaft.
3. Position the fixing ring. The notches in the fixing ring must always be pointing towards the
saw (see fig. 2). These must also be positioned precisely on the spigots.
4. Screw the retaining screw in manually until hand tight.
5. Then tighten firmly using the fork wrench, as described above.
The tightening torque is 6 - 7 Nm. Caution! Do not overtighten.
22
Fig. 1:
Releasing the retaining screw
Inserting the saw blade
Oscillo-Saw TITANIUM II
Fig. 2:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hb 8884-aHb 8884-bHb 8884-gHb 8884-sHb 8884-v

Tabla de contenido