Inserting The New Saw Blade; Operation - HEBU medical GOLD II Autopsie Instrucciones Para El Uso Y El Servicio

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

EN

5.2 Inserting the new saw blade

1. Check that the female thread and saw blade support of the drive shaft are clean.
2. Position the new saw blade.
It makes no difference which side of the saw blade is facing inwards or outwards.
Important!
The notches in the saw blade must be positioned precisely flush on the spigots of the
drive shaft.
3. Position the fixing ring. The notches in the fixing ring must always be pointing towards
the saw (see fig. 2). These must also be positioned precisely on the spigots.
4. Screw the retaining screw in manually until hand tight.
5. Then tighten firmly using the fork wrench, as described above.
The tightening torque is 6 – 7 Nm. Caution! Do not overtighten.

6 Operation

Switching on and off (see fig. 3)
Switching on:
Push the switch forwards
Switching off:
Push the switch back
Setting the oscillating frequency (see fig. 4)
Select the oscillating frequency when the motor is running.
The optimum oscillating frequency can be selected at the s
peed gauge depending on the properties of the plaster cast.
Oscillating frequency preselection using the electronic
speed gauge.
Lowest oscillating frequency
Highest oscillating frequency
24
Inserting the saw blade
( I )
( 0 )
( 1 )
( 8 )
Oscillo-Saw GOLD II Autopsy
Fig. 2:
O
I
Fig. 3
Fig. 4

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Hb 8891-aHb 8891-bHb 8891-gHb 8891-sHb 8891-v

Tabla de contenido