2.Disinfection
After each use, it is necessary to disinfect the components of the cup,
including medication cup,lid, spray assembly,mask or mouthpiece., etc.,
as follows:
1) Disinfection with hydrogen peroxide
Disinfect all the components by placing them in 3% hydrogen peroxide
for 10 minutes, including cup components (including medication cup,lid),
spray assembly,mask or mouthpiece. Or wipe and disinfect the medicine
cup component (including medication cup,lid), spray assembly,mask or
mouthpiece.After disinfection, rinse all the parts with distilled water,then
wipe with clean and sterile medical gauze or air-dry naturally to keep all
parts dry.
A. Please study the user guidance of hydrogen peroxide and do not
immerse in solution for a long time.
B. Do not use strong oxidizing agents such as perchlorate or
disinfectants that are corrosive to metals, polymer compounds or
polymers.
2) Ethanol disinfection
Place the liquid cup components(including medication cup,lid), spray
assembly,mask or mouthpiece in 75% medical ethanol for 10 minutes for
disinfection. Or wipe and disinfect the medicine cup component(including
medication cup,lid), spray assembly,mask or mouthpiece. After
disinfection, rinse all the parts with distilled water,then wipe with clean
and sterile medical gauze or air-dry naturally to keep all parts dry.
Notes:
•Disinfectants remaining on components need to be wiped with
sterile medical gauze to ensure safe use next time. Do not touch the
central area of the mesh diaphragm when washing or cleaning the spray
connector,so as to avoid damaging the mesh diaphragm.
3.Drying
• Dry the parts carefully using a soft cloth.
• Shake the spray assembly gently from side to side (5 – 10 times), so
that the water inside the mesh is removed from the tiny holes.
• Place the individual parts on a clean, dry and absorbent surface
and leave them to dry completely (at least 4 hours).Note Please ensure
that the parts are completely dry after cleaning, otherwise the risk of
bacterial growth is increased.Put the parts together again if they are