Cleaning And Disinfection Method - Feellife Air Angel Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Keep quiet and relaxed in the course of treatment.
3) Liquid will be coagulated around the spray assembly and mesh
disc, which will affect the nebulization effect after nebulization. It should
stop nebulization, remove the nozzle and other accessories. Then use
a clean medical gauze to wipe the residue. Do not touch the mesh disc
center spray area to prevent damage to the mesh disc.
4) Essential oils, cough syrups, gargling solutions and drops to be
used as a rub or in a steam bath are wholly unsuitable for inhalation
using a nebulizer. These additives are often viscous and can impair the
correct functioning of the device and therefore the effectiveness of the
application in the long term.

5. Cleaning and Disinfection Method

After each use, it is necessary to clean and disinfect the cup
components (including medication cup,lid), spray assembly,mask or
mouthpiece.The specific methods of cleaning and disinfection are
recommended as follows:
1.Cleaning
Please turn off the power when cleaning machine. Do not connect the
machine with power supply.
1) Remove the components from the main unit: the medicine cup
components (including medication cup,lid), spray assembly,mask or
mouthpiece.; and soak all the components in clean warm water (which is
no more than 40 degrees centigrade) for about 5 minutes.
2) After cleaning, wipe all the components with clean and sterile
medical gauze, and keep them dry sufficiently.
3) Wipe the outer shell of main unit. If there is residue medicine
remaining at the electrode contact, please clean it with wet sterile
medical gauze. After cleaning, keep the main unit dry.
4) Store the all parts in a dry and clean place to avoid contamination.
Notes:
• The main unit can not be washed with water to prevent water from
entering the main unit.
• Use clean sterile gauze to wipe off water on the main unit and
components and keep them dry to ensure safe use next time.
•The masks must not be placed in hot water!
14

Publicidad

Tabla de contenido
loading

Tabla de contenido