When multiple troubleshooting alarms, they will show alternately. Alarms and indicators
depend on the highest priority.
The above alarms are technical alarms.
Within 3 minutes after power is on, the oxygen concentration of less than 82% will not alarm.
When a troubleshooting alarm occurs, identify the type of failure that is displayed on the LCD
and contact your service provider immediately.
The recommended test interval of the alarm system is 18 months.
The alarm sound pressure level is of 60~80 decibels

Troubleshooting

Problem
Power on, the
The start capacity of the
equipment is not
compressor is broken or the
working.
compressor is not working.
Power on, the
The power cord may not be
equipment is not
plugged correctly or there may be
working, or works
un-continuously.
No oxygen outlet or
* Oxygen cannula kinked or
the outlet flow is
too low.
* The humidifier bottle is not
The ball in flow
* The flow knob is not tighten.
meter is
* The knob was turned abruptly
uncontrolled by the
knob.
Water goes back to
* There is a temperature
the cannula.
difference caused by the
* The device is too close to the
wall, curtains or furniture.
* The location of the equipment
and cannula is different.
DOMU119_EN-FR-ES-RU-DE_v4_20230105_ Horizon® S5
Why it happened
a bad contact.
Call your service provider, if you still cannot solve
blocked.
Call your service provider, if you still cannot solve
tighten.
or too fast.
weather.
* Keep the equipment and cannula in the same
What to do
Call your service provider or dealer.
* Check if the power cord is damaged.
* Check if the power cord is plugged correctly.
the problem.
* Unknit the cannula.
* Re-install the humidifier lid.
the problem.
* To tight the knob, turn the knob slowly and
softly.
Call service provider, if still cannot solve the
trouble.
* Dry the inside of humidifier lid.
* Do NOT use hot water.
* Do NOT over-fill the humidifier.
room to maintain the same temperature.