Table 201 - Guidance and MANUFACTURER'S declaration - ELECTROMAGNETIC EMISSIONS - for all equipment and systems
Guidance and manufacturer's declaration - electromagnetic emissions
The Radii Plus is intended for use in the electromagnetic environment specified below. The customer or the user of the Radii Plus should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker emissions IEC 61000-3-3
Table 202 - Guidance and MANUFACTURER'S declaration - Electromagnetic IMMUNITY - for all equipment and systems
Guidance and manufacturer's declaration - electromagnetic immunity
The Radii Plus is intended for use in the electromagnetic environment specified below. The customer or the user of the Radii Plus should assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge (ESD) IEC 61000-4-2
Power frequency (50/60 Hz) magnetic field
IEC 61000-4-8
Table 204 - Guidance and MANUFACTURER'S declaration - Electromagnetic IMMUNITY - for all equipment and systems that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The Radii Plus is intended for use in the electromagnetic environment specified below. The customer or the user of the Radii Plus should assure that it is used in such an environment.
IMMUNITY test
Radiated RF IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the Radii Plus is used exceeds the applicable RF compliance level above, the Radii Plus should be observed to verify normal operation. If abnormal performance is observed, ad-
ditional measures may be necessary, such as re-orienting or re-locating the Radii Plus.
Table 206 - Recommended separation distances between portable and mobile RF communications equipment and the equipment and systems - for equipment and systems that are not
LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the Radii Plus
The Radii Plus is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Radii Plus can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Radii Plus as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output power of trans-
mitter W
0,01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Compliance
Group 1
Class B
Not applicable
Not applicable
IEC 60601 test level
Compliance level
+/- 6 kV contact
+/- 6 kV contact
+/- 8 kV air
+/- 8 kV air
3 A/m
3 A/m
IEC 60601 test level
Compliance level
3 V/m 80 MHz to 2,5 GHz
3 V/m
Separation distance according to frequency of transmitter m
80 MHz 10 800 MHz
0.12
0.38
1.2
3.8
12
Electromagnetic environment - guidance
The Radii Plus uses RF energy only for its internal function. There-
fore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
The Radii Plus is suitable for use in all establishments, including
domestic establishments. The Radii Plus is powered by 2 x 4.2V
DC Lithium ion batteries thus Harmonic emission and Voltage
fluctuation/flicker emissions are not applicable.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be at
least 30%.
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital environ-
ment.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the TENS PRO 900, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance
level in each frequency range .b
Interference may occur in the vicinity of equipment marked with
the following symbol:
800 MHz a 2,5 GHz
0.23
0.73
2.3
7.3
23