Safety and performance
Declaration – electromagnetic immunity
Withings Body Comp is intended for use in the electromagnetic environment specified below.
The customer or the user of the Withings Body Comp should ensure that it is used in such an environment.
Immunity test
IEC 60601 Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150kHz to 80MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Electrostatic discharge (ESD)
Contact: ±8 kV
IEC 61000-4-2
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Power frequency (50/60Hz)
30 A/m
magnetic field IEC 61000-4-8
50 Hz or 60 Hz
Compliance Level
Electromagnetic environment – guidance
N/A
N/A
N/A
Portable and mobile RF communications equipment should be used
no closer to any part of the equipment or system including cables than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter. Interference may occur in the vicinity
of equipment marked with the following symbol:
Contact: ±8 kV
Floors should be wood, concrete, or ceramic tile. If floors are covered
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
with synthetic material the relative humidity should be at least 30%.
30 A/m
The power frequency magnetic field should be at levels characteristic of
50 Hz or 60 Hz
a typical location in a typical commercial or hospital environment.