Model SH-200D-S and SH-200D-S-L have a
see-saw rocking action that creates a wave
motion within vessels such as culture flasks, Petri
dishes etc. Speed range is 5 to 70 rpm. Max load
3kg and 10kg respectively.
Models SH-200D-M and SH-200D-M-L
are designed for use with microtitre plates
and microcentrifuge tubes. High speed
combined with a tiny orbit creates the vibrational
shaking action required for mixing very small
volumes. Speed range is 250 to 1,200 rpm.
IMPORTANT Maximum load is 1kg.
4. Preparation For Use –
All Models
4.1 Electrical Installation
THIS INSTRUMENT MUST BE
EARTHED
Before connection please read and understand
these instructions and ensure that the line supply
corresponds to that shown on the rating plate.
120V, 60Hz, ~, single phase
230V, 50Hz, ~, single phase
The power consumption of each unit is 50W.
The instruments are fitted with an IEC socket at
the rear of the instrument for connection of the
mains lead.
Caution: Fuses fitted in both live and neutral
lines.
These units are supplied with two mains leads
fitted with IEC plugs for connection to the
instrument. One lead has a U.K. 3 pin plug and
the other has a 2-pin "Shuko" plug for
connection to the mains. Choose the lead
appropriate for your electrical installation and
discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative. This involves cutting off the
moulded plug, preparing the cable and
connecting to the rewireable plug in accordance
with its instructions.
IT IS IMPORTANT THAT THIS OPERATION
SHOULD ONLY BE UNDERTAKEN BY A
QUALIFIED ELECTRICIAN
NOTE: Refer to the equipment's rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated.
The wires in the mains cable are coloured as
follows:
BROWN or BLACK
BLUE or WHITE
GREEN/YELLOW
The appropriate mains lead should be connected
to the instrument BEFORE connection to the
mains supply.
Should the mains lead need replacement a cable
2
of 1mm
of harmonised code H05VV-F
connected to an IEC320 plug should be used.
N.B. The U.K. mains lead is protected by a 10A
fuse mounted in the plug top.
IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN
4.2 Connect to the electricity supply – DO NOT
SWITCH ON.
4.3 Place the unit on a firm level non slip
surface ensuring that there is sufficient free
space on all sides to allow movement without
coming into contact with anything during use.
4.4 Assemble cradle and rollers onto platform
where applicable, making sure that the thumb
screws have been tightened prior to operation.
4.5 Load the shaker/rocker with samples
ensuring that the load is evenly balanced and
does not exceed the maximum load stated.
4.6 Adjusting the angle of tilt on the 3D gyratory
motion rockers, models SH-200D-G and
SH-200D-G-L. Set the rocker to the minimum
speed (5rpm). Underneath the platform you will
see a cylindrical shaft with a nut on it, stop the
rocker when the nut faces you. Do not loosen
the nut. The platform can now be tilted at right
angles to the front fascia. Once you are happy
with the angle of tilt restart the rocker.
- LIVE
- NEUTRAL
- EARTH