CONFIGURATION MENU
8. The calibration of valve anticipation factor is performed by
dispensing 0.26 gal (1 liter) at the working flow rate. This
preset configuration will already appear on the display:
MANUAL SET OF CALIBRATION FACTOR
1. Simultaneously press the "ON" and "SELECT" buttons
until the setup menu is displayed..
3. Press and hold "SELECT" until "Cfactr" flashes in
second line and the current calibration factor is
displayed.
5. Confirm the selection by pressing the "ON" button, the
display returns to the main menu. Pressing the "ON"
button again exits the configuration menu.
Calibration factor calculation
Manual calculation of calibration factor can be
made with following equation:
quantity shown on tdisplay
=
Factor
actual quantity dispensed
Balcrank™ Corporation · Weaverville, NC 28787 · 800-747-5300 · 800-763 0840 Fax · www.balcrank.com
Fig. 14
Fig. 15
Fig. 17
9. Press AUTO and then start dispensing. When the preset
0-26 gal (1 liter) is reached, the gun will automatically
calibrate the anticipation by displaying the new value on
the screen. The display shows "done" and automatically
returns to the main menu.
2. Press "SELECT" until the next menu is displayed.
4. Set the value of the new calibration factor. Pressing the
"0.1" quantity button increments the units and the "10"
button decrements the units.
x 1000
Fig. 16